ClearanceBrief
Recall Z-1281-2019 ยท JAK System, X-Ray, Tomography, Computed

Neusoft Medical Systems Co., Ltd.

NeuViz 16 Essence Multi-slice CT Scanner System Product Usage: The Multi-Slice CT Scanner System can be used as a whole body computed tomography x-ray system featuring a continuously rotating x-ray tube and detector array. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body from either the same axial plane taken at different angles or spiral p

DateOct 22, 2018
StatusTerminated
ReasonSoftware defect: For Helical scan with ClearView function, when small arc (uninterrupted scan) occurs during scan, there will be some electromagnetic interference in the system which may affect the control signal of the flying focal spot, resulting in an incorrect flying focal spot sequence. If the system does not receive error, it would continue to complete the scan with the incorrect flying focal spot sequence, having an effect on the reconstruction of some images, resulting in image loss.
Root causeSoftware design
ActionA customer notification letter will be sent to affected customers by email. All the affected systems will be corrected with the upgraded software in the field free of charge.
Quantity2
DistributionUS Nationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171198NeuViz 16 Essence Multi-slice CT Scanner System
Neusoft Medical Systems Co., Ltd., Dec 7, 2017