ClearanceBrief
Recall Z-1701-2019 ยท NEY System, Ablation, Microwave And Accessories

Angiodynamics Inc. (Navilyst Medical Inc.)

AngioDynamics Solero Microwave Tissue Ablation (MTA) System: (1) Solero Generator Refurbished PG Catalog Number: 12740000D UPN: H78712740000D0 (2) Solero Generator Refurbished US PG Catalog Number: 12740000USD ;UPN: H78712740000USD0

DateApr 17, 2019
StatusTerminated
ReasonUpdated Warning Statements in the Operator's Manual/Directions for Use and Software Updates to reduce the risk of potential patient harm due to a delay in performing a procedure
Root causeDevice Design
ActionAngioDynamics issued on April 17, 2019 AngioDynamics initiated a Field Correction by issuing notifications via Federal Express. Consignees are instructed to replace the existing Solero Generator User Manual with the revised version provided with the Notification. Once the replacement has been made, the consignees are instructed to complete and return the Reply Verification Tracking Form, provided in the notification. AngioDynamics will also install a software update to help reduce the incidence of Error 0001 which can occur during initial start-up. This improvement will be implemented on the n
Quantity178 units
DistributionWorldwide distribution - US Nationwide distribution and countries of AE AT AU BE CA CH CL CO CZ DE DK DZ EG ES FR GB GE GR HK IE IL IN IQ IT LB LT MC MT MX NL NZ PL PT SG TH TR TW VN ZA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162449Solero MTA System, Solero Microwave Tissue Ablation Applicator (14cm), Solero Microwave Tissue Ablation Applicator (19cm
AngioDynamics, Inc., May 5, 2017