Recall Z-1850-2019 ยท KQO Automated Urinalysis System

Arkray Factory USA, Inc.

AUTION HYBRID AU-4050

DateMay 3, 2019
StatusTerminated
ReasonThis correction is being initiated due to a software issue which results in the possibility of incorrect patient information being assigned to sample results when the following rare combination of three specific events were to occur: 1) No measurement results are generated due to an error by a urine sediment measurement. 2) The instrument is shutdown incorrectly. 3) An item rack is used for subsequent sample measurements.
Root causeSoftware design
ActionThe firm initiated the field correction by email on 05/03/2019. The letter stated how to recognize a product failure and the actions to be taking by the customer until the software upgrade is ready for installation.
Quantity159 units
DistributionUS
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121456AUTION HYBRID AU-4050 FULLY AUTOMATED INTEGRATED URINE ANALYZER SYSTEM; URIFLET S 9HA URINE TEST STRIPS; AUTION CONTROL
Arkray, Inc., Feb 1, 2013