Recall Z-1699-2019 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Renal Therapies Group, LLC

2008T BlueStar Machine Software v.2.72-Hemodialysis machine Model Number: 191126 - Product Usage: The 2008T hemodialysis machine is indicated for acute and chronic dialysis therapy in a healthcare facility.

DateApr 3, 2019
StatusTerminated
ReasonPower Logic Board Fail Message on the 2008T BlueStar Machine with software version 2.72.
Root causeNonconforming Material/Component
ActionFresenius issued An IMPORTANT MEDICAL DEVICE CORRECTION notification explains the problem, health risk and that a Fresenius Service Technician (FST) will be contacting them to schedule the correction. Questions, please contact Fresenius Technical Services at 800-227-2572.
Quantity257 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1739722008T BlueStar Hemodialysis Machine