ClearanceBrief
Recall Z-1692-2019 ยท DSI Detector And Alarm, Arrhythmia

Philips North America, LLC

IntelliVue MX40 Patient Monitor, Product number 865350, Service # USA: 453564262491 and 453564262511 - Product Usage: Used primarily as a traditional telemetry medical device. It connects to the IntelliVue Information Center by way of a wireless network. When the MX40 is connected to the IntelliVue Information Center the IntelliVue Information Center provides the primary patient monitoring and ala

DateMar 28, 2019
StatusOpen, Classified
ReasonThe MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .
Root causeSoftware design
ActionOn March 28, 2019, Philips Healthcare distributed URGENT MEDICAL DEVICE RECALL notices (FSN 86201863A) dated March 22, 2019 to customers via courier service. Actions To Be Taken by Customer/User at the time was as follows: - MX4O monitors can continue to be used while awaiting a software upgrade. - Customers should maintain an adequate supply of fully charged batteries given the approximately 25% increased power consumption. Switch to continuous or manual SO2 measurements if your unit is using automatic measurement mode. On April 24, 2019, Philips Healthcare distributed UPDATE URGENT ME
DistributionWorldwide distribution - US Nationwide and to 71 countries.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113125INTELLIVUE MX40 PATIENT MONITOR
Philips Medical Systems, Nov 17, 2011