ClearanceBrief
Recall Z-1408-2019 ยท JSG Culture Media, Non-Selective And Non-Differential

Remel Inc

Thio Med w/o Ind, w/Dex (18ml) 100/PK Product Usage: Recommended for use in qualitative procedures as a general purpose media for the cultivation of aerobes and anaerobes

DateApr 17, 2019
StatusTerminated
ReasonPotential for products to fail performance testing per the Instructions for Use (IFU) starting at 106 days after date of manufacturing. Expiration date is shortened from 9 months to 105 days.
Root causeUnder Investigation by firm
ActionThermo Fisher Scientific/ Remel notified consignees on about 04/17/2019 via " URGENT MEDICAL DEVICE FIELD ACTION RECALL" letter. Instructions included to locate all affected in inventory, destroy any affected product identified, contact Customer Services or local distributor to arrange for replacements, notify all customers if product was further distributed, and complete and return the Field Action Response form. Customers were also instructed that the requirement for review of reported test results using product beyond the updated expiry date should be determined by the appropriate techni
Quantity76 units
DistributionWorldwide Distribution -US Nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico. International distribution to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.