ClearanceBrief
Recall Z-1394-2019 ยท GEX Powered Laser Surgical Instrument

Boston Scientific Corporation

AURIGA XL 4007 GENERAL - ZERO COST SYSTE UPN: FS4007GZ0 Product Usage: The Auriga XL 4007 Laser System and the Auriga 30 Laser System including a fiber optic delivery system is intended to be used in surgical procedures such as open, laparoscopic and endoscopic, to perform incision, excision, resection, ablation, vaporization, coagulation and hemostasis of soft tissue in medical specialties in

DateMay 22, 2019 (initiated Apr 8, 2019)
StatusTerminated
ReasonCertain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.
Root causeComponent design/selection
ActionBoston Scientific issued to the user level the correction/removal via a customer notification letter via FedEx or hand delivery by a Boston Scientific agent on 4/9/19. Letter provides recommendations to the user until a field repair can be completed or a replacement console can be provided. A Boston Scientific representative will be in contact with you directly to schedule a field repair or replacement at your facility. The console can continue to be used during this time following the recommendations within the letter.
Quantity6 (OUS)
DistributionWorldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111475AURIGA XL
Starmedtec GmbH, Sep 28, 2011