Recall Z-1324-2019 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America, LLC

SureSigns VS4 Government Bundle, Product Number 863286. For monitoring the physiological parameters of patients (NBP, SpO2 & Temperature).

DateApr 19, 2019
StatusOpen, Classified
ReasonSystem software inhibits the monitor and as a result does not measure, display and alarm for pulse rates above 240 beats per minute.
Root causeSoftware design
ActionOn April 19, 2019 Philips Healthcare issued URGENT MEDICAL DEVICE CORRECTION notices to customers via courier service. Actions to be taken by customer: Upon receipt of this notification, promptly perform the software upgrade to each of your SureSigns VS3/VS4 monitors, as called out in the attached Pulse Rate Software-Installation Instructions. It details how to obtain and install the system software update. It also provides instructions on how to export the file that is created during software installation and instructions where to email the file back to Philips. If you need any further
Quantity366
DistributionWorldwide distribution. US Nationwide including Puerto Rico, Andorra, Argentina, Aruba, Australia, Austria, Bahrain, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Dominican Republic, Egypt, France, French Guiana, French Polynesia, Germany, Gibraltar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163649SureSigns VS3; SureSigns VS4
Philips Medical Systems, Mar 31, 2017