Recall Z-1754-2019 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare, LLC

Precision 500D The Precision 500D R&F X-ray System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations.

DateJun 10, 2019 (initiated Mar 29, 2019)
StatusTerminated
ReasonThere has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
Root causeDevice Design
ActionOn March 29, 2019, GE Healthcare sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated March 29, 2019 to its consignees/customers. The letter described the product, problem and actions to be taken. Customers were instructed to do the following: Please follow the precautions below until the monitor feet on your unit are replaced: -The CRT monitor and suspension should be inspected and used with caution. Make sure the monitors are secured properly to the tray before each use. -Immediately stop using the system if any looseness of the monitor is noticed and contact your GE service represent
Quantity825 (38 US and 787 OUS) in total
DistributionWorldwide Distribution: US (nationwide) to states of: AL, CA, CT, FL, GA, HI, IL, LA, MD, MI, MN, MO, NH, NY, OK, PA, TN, and VA; and countries of: Algeria, Argentina, Belgium, Bosnia & Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Egypt , France, F
FDA recordOpen on fda.gov