Recall Z-1696-2019 ยท IYE Accelerator, Linear, Medical

Accuray Incorporated

CyberKnife M6, Part Numbers: a) 0660000: CyberKnife Treatment Delivery System b) 053101-010: CyberKnife M6 FI Treatment Delivery System c) 053201-010: CyberKnife M6 FM Treatment Delivery System d) 053301-010: CyberKnife M6 FIM Treatment Delivery System e) 053401-010: CyberKnife M6 FI+ Treatment Delivery System f) 053501-010: CyberKnife Matrix Tier 2 TDS

DateApr 10, 2018
StatusTerminated
ReasonThere is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.
Root causeOther
ActionThe firm issued a field safety notice by letter on 04/10/2019. The letter explained the issue and provided safety instructions. All impacted customers with affected systems will be contacted to arrange for a software upgrade.
Quantity68 units
DistributionUS & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150873CyberKnife M6 FI Sytem, CyberKnife M6 FM System, CyberKnife M6 FIM System
Accuray Incorporated, Jul 1, 2015