ClearanceBrief
Recall Z-1927-2019 ยท JQP Calculator/Data Processing Module, For Clinical Use

Roche Diagnostics Corporation

cobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.

DateJul 3, 2019 (initiated Apr 4, 2019)
StatusTerminated
ReasonThere is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
Root causeNonconforming Material/Component
ActionOn April 4, 2019, Roche issued Urgent Medical Device Correction notices (UMDC TP-00620) to customers via UPS Ground (receipt required). Customers were advised to take the following actions: 1) Inspect the tray input flap on the cobas p 501 or p 701 post-analytical unit to determine if it is loose. If it is not fastened securely or becomes loose prior to the part being replaced, contact your Roche Field Engineer Specialist (FES) to repair the instrument. 2) Complete and return the fax form (TP-00621) via fax or email 3) File the UMDC for future reference. Customers with questions may cont
Quantity4 units
DistributionUS Nationwide Distribution in the states of AL, CA, FL, IA, IL, KY, MD, OR & PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.