Recall Z-1774-2019 ยท EZW Stimulator, Electrical, Implantable, For Incontinence

Medtronic Neuromodulation

InterStim(TM) System, Model Numbers: a) TH90G01 b) TH90GFA c) TH90G02 d) TH90G03 Product Usage: The Medtronic Model A510 Clinician application (app) is intended for use with the HH90 Handset and TM90 Communicator to program, adjust, and troubleshoot the Medtronic Models 3023 and 3058 InterStim neurostimulators for sacral neuromodulation therapy. The clinician uses the Clinician app to program set

DateJun 7, 2019 (initiated May 15, 2019)
StatusOpen, Classified
ReasonThere is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).
Root causeSoftware design
ActionMedtronic sent an Urgent Medical Device Safety Notification letter dated May 2019, to US Physicians and European physician. The notifications were delivered by mail, personal delivery by Medtronic Representatives, fax, or equivalent method. A confirmation form will be used to document receipt and understanding of the notification, and a minimum three attempts will be made to obtain confirmation from non-responding physicians that the notification has been received and understood.
Quantity13979 units
DistributionWorldwide Distribution - US Nationwide & PR, and Germany, Switzerland, Italy, Spain, France, UK, Norway, Denmark, Finland, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.