Recall Z-1295-2019 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Cios Alpha with software version VA30-mobile C-arm x-ray Material # 11105200 Product Usage: The Cios Alpha is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurolo

DateMar 22, 2019
StatusTerminated
ReasonContinuous operation of the Cios Alpha at high tube outputs may result in increased wear of radiation emitting components, risk of premature failure of the Monoblock and loss of imaging X-rays exists
Root causeDevice Design
ActionSiemens issued Customer Safety Advisory Letter dated 3/22/19 advising of problem, health risk and action to take: recommends to avoid overwriting the organ programs, as this can lead to frequent operations. Siemens is distributing this letter and the addendum to the user manual to all potentially affected users Contact our service organization for any additional questions at 1-800-888-7436.
Quantity2 US
DistributionUS Nationwide Distribution in the states of NC, MA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181560Cios Alpha
Siemens Medical Systems USA, Inc., Oct 24, 2018