Recall Z-1240-2019 ยท LNH System, Nuclear Magnetic Resonance Imaging
Philips Medical Systems Nederlands
Ingenia Ambition S (781359), Ingenia Ambition X (781356) with Magnet Energization Device (MED), Philips MR systems are indicated for use as a diagnostic device. The purpose of the MED is to supply power to charge the magnet. MED is abbreviation for Magnet Energize Device
| Date | Mar 25, 2019 |
| Status | Terminated |
| Reason | Magnet Energization Device (MED) can have a residual voltage on output terminal, may cause electrical shocks to personnel servicing the MED, MDD or connected cabling. |
| Root cause | Device Design |
| Action | Philips issued an "URGENT - Medical Device Correction" letter on 3/25/2019 via certified mail to its customers. Philips Key Markets are responsible for distributing the letters outside of the U.S. Letter identifies the problem, health risk and actions to be taken. The customers were informed that no action from the customer or user of the device is required, as the Magnet Energization Device (MED) is located in a cabinet in the Technical Room, with the output terminals of the MED located behind locked protective covers. Only qualified servicing personnel have access to this. the Magnet PHILIPS |
| Quantity | 4 units |
| Distribution | US Distribution to state of: FL and Internationally to: Switzerland and Germany. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K181479 | Ingenia Ambition S and Ingenia Ambition X Philips Medical Systems Nederland B.V., Aug 3, 2018 |