ClearanceBrief
Recall Z-0165-2020 ยท FSA Lift, Patient, Non-Ac-Powered

Community Products, LLC

Rifton TRAM Low Base, SKU K320, powered lift device

DateMar 21, 2019
StatusTerminated
ReasonA design change to the load-bearing quick release buckle renders the springs in the latch mechanism susceptible to fatigue failure after repeated use.
Root causeNonconforming Material/Component
ActionUrgent Medical Device Correction emails were sent to customers beginning on March 21, 2019. Customers were contacted via telephone beginning April 5, 2019.
Quantity1079
DistributionThe products were distributed US nationwide. The products were distributed to the following foreign countries: Foreign Countries: Worldwide, including Canada, United Kingdom, Germany, Japan, Singapore, South Korea, United Arab Emirates, Taiwan, Denmark, Italy, Hong Kong, Ireland, Australia, New Ze
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.