Recall Z-1265-2019 ยท JAK System, X-Ray, Tomography, Computed
GE Healthcare, LLC
NM/CT 870 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can
| Date | Mar 20, 2019 |
| Status | Terminated |
| Reason | During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy. |
| Root cause | Device Design |
| Action | GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity | 27 units |
| Distribution | US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K093514 | DISCOVERY MODEL NM/CT 670 Ge Medical Systems F.I. Haifa, Dec 10, 2009 |