Recall Z-1263-2019 ยท JAK System, X-Ray, Tomography, Computed
GE Healthcare, LLC
NM/CT 860 Product Usage: The GE NM/CT 860 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and ana
| Date | Mar 20, 2019 |
| Status | Terminated |
| Reason | During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy. |
| Root cause | Device Design |
| Action | GE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity | 6 units |
| Distribution | US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K173816 | NM/CT 850, NM/CT 860 GE Healthcare, Apr 17, 2018 |