ClearanceBrief
Recall Z-1261-2019 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

NM 830 Product Usage: The Discovery NM 630 is an all-purpose dual detector nuclear imaging system intended for general nuclear medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and imaging features designed to enhance image quality. The main components of the Discovery NM 630 syste

DateMar 20, 2019
StatusTerminated
ReasonDuring installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Root causeDevice Design
ActionGE Healthcare sent an Urgent Medical Device Correction letter dated March 20, 2019 to affected customers. The letter identified the affected product, problem and actions to be taken. GE Healthcare will inspect and, if required, correct all affected products at no cost to you to ensure that all design redundancies are in place. A GE Healthcare representative will contact you to arrange for the correction. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity37 units
DistributionUS and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173816NM/CT 850, NM/CT 860
GE Healthcare, Apr 17, 2018