Recall Z-1364-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Ortho-Clinical Diagnostics
VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest
| Date | Mar 20, 2019 |
| Status | Terminated |
| Reason | Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported |
| Root cause | Software design |
| Action | Ortho Clinical Diagnostics issued on 20 March 2019, a customer letter via FedEx overnight courier and/ or ORTHO PLUS e-Communications and and/or US Postal Service Priority Mail (for PO Boxes only) customers who were shipped affected VITROS XT 7600 Systems. Software Version 3.4.2 (MOD 4) will be provided for download or DVD format. Contact Ortho Care" Technical Solutions Center at 1-800-421-3311. |
| Quantity | 85 units: US=19;OUS=66 |
| Distribution | Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K182063 | VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides, VITROS Chemistry Products TBIL Slides, VIT Ortho-Clinical Diagnostics, Inc., Oct 30, 2018 |