ClearanceBrief
Recall Z-1721-2019 ยท MED Sterilant, Medical Devices

Kerr/Pentron, DBA Kerr Corporation and Pentron Clinical

Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution

DateJul 5, 2019 (initiated Jul 13, 2018)
StatusTerminated
ReasonImmersion time stated on the bottle is 45 minutes while the product instructions say 90 minutes.
Root causeError in labeling
ActionOn July 13, 2018, Metrex Research Corp., a subsidiary of Kerr Corporation, sent URGENT NOTICE: Regimen Glutaraldehyde letters to their consignees instructing them to check their inventory and complete the enclosed Field Action Acknowledement and Product Return Form. Distributors were requested to conduct sub-recalls to their consignees.
Quantity31904
DistributionAZ, CA, CO, FL, GA, IL, KY, LA, MA, MN, MO, NJ, NY, OH, PA, SC, TN, TX, UT, VA, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K931052METRICIDE 28
Metrex Research Corp., Dec 6, 1994