Recall Z-1213-2019 ยท HDA Forceps, Obstetrical

COVIDIEN LLC

Procedure kits include devices from the affected lots distributed OUS only: (1)BOX KIT00529H KIT COLECISTECTOMIA ACCESO (2)BOX KIT00680 KIT COLECISTECTOMIA- (3)BOX KIT00681 KIT APENDICECTOMIA (4)BOX KIT00897 KIT COLON IZQUIERDO (5)BOX LAPL1 ZESTAW DO CHOLECYSTEKTOMII (6)BOX PST01134 LAP CHOLE KIT ST JOHN S (7)KIT00599 LAPARO GENERICO X1 (8)KIT00608 APENDICECTOMIA LAP X1 (9)KIT2176CV BORNHOL

DateMar 13, 2019
StatusTerminated
ReasonSterilization method used was not consistent with the labeling and the approved sterilization method
Root causeProcess control
ActionMedtronic notified consignees via an "URGENT MEDICAL DEVICE RECALL" letter by Federal Express and/or certified mail on March 13, 2019. The letter informs customers of the recall of twenty-one production lot of Covidien Endo Clinch" and Endo Grasp" Auto Suture" graspers 5mm as the sterilization method used for these lots was not consistent with the labeling and the approved sterilization method. Customers are instructed to take the following actions: 1. Please immediately quarantine and discontinue use of the affected item codes and lots listed above. 2. Please return affected product as indic
Quantity148 units
DistributionWorldwide Distribution: US (Nationwide) and to countries of: Canada, Chile, Costa Rica, Guatemala, Japan, Mexico, Panama, Slovakia, Spain, Switzerland and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K903205AUTO SUTURE(R) ENDOSCOPIC CLAMP
United States Surgical, A Division of Tyco Healthc, Oct 11, 1990