Recall Z-1297-2019 ยท IYE Accelerator, Linear, Medical

Elekta Limited

Elekta Unity systems Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Info

DateMay 8, 2019 (initiated Apr 3, 2019)
StatusTerminated
ReasonThere is no warning in the Elekta Unity manual for the administration of gadolinium-based or other contrast agents for imaging or image guidance purposes has not been validated.
Root causeLabeling design
ActionElekta notified customers on about 04/03/2019 via email with "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" letter. Instructions included to ensure all staff working with Elekta Unity are aware of the issue and recall information and post the recall notice in a place accessible to all users, e.g. Instructions for use, until the recall is closed. Additionally customer were asked to complete and return the Acknowledgement Form.
Quantity9 devices
DistributionWorldwide Distribution - US Nationwide in the states of Texas and Wisconsin. Distribution internationally to Denmark, England, Germany, Italy, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K182076Elekta Unity
Elekta Limited, Dec 4, 2018