ClearanceBrief
Recall Z-1230-2019 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

OriGen Biomedical, Inc.

ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV13F

DateJun 4, 2019 (initiated Jan 17, 2017)
StatusTerminated
ReasonClamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Root causeProcess control
ActionOn 01/17/2017, Technical Bulletin 17.01 was mailed to customers. Customers were informed to clamp all sizes of catheters only in the middle. The following contact information was provided: US: 1-512-474-7278, and Germany: +49-7475-915591.
Quantity145
DistributionUS: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113869ORIGEN REINFORCED DUAL LUMEN CATHETER
Origen Biomedical, Inc., Jul 26, 2012