Recall Z-1235-2019 · JZO System, Test, Thyroid Autoantibody
Siemens Healthcare Diagnostics, Inc
Siemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 10995461
| Date | Dec 7, 2018 |
| Status | Terminated |
| Reason | Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU). |
| Root cause | Nonconforming Material/Component |
| Action | Siemens issued Urgent Medical Device Corrections, to Siemens Healthcare Diagnostics customers in the United States on December 7, 2018 delivered to customers on December 10, 2018. Urgent Field Safety Notices sent out to the country organizations on December 7, 2018 for distribution to customers outside the United States. These letters inform customers that Siemens Healthcare Diagnostics identified a bias with the affected lots of ADVIA Centaur aTG and Atellica IM aTG and action to take and resolution of the issue with new lots of the reagents. Questions contact your Siemens Customer Care Cen |
| Quantity | 1489 kits |
| Distribution | Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K012777 | ACS: 180 AND ADVIA CENTAUR ANTI-THYROGLOBULIN IMMUNOASSAYS Bayer Corp., Oct 22, 2001 |