ClearanceBrief
Recall Z-1200-2019 ยท IZI System, X-Ray, Angiographic

GE Healthcare, LLC

Innova IGS 630, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.

DateMar 15, 2019
StatusOpen, Classified
ReasonThere is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
Root causeDevice Design
ActionGE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customer were informed that they can continue to use their Innova IGS 6 or Innova IGS 630 Systems but must follow the following safety instructions: 1. Do not change the size of the field of view from 30cm to 20cm or 20cm to 30cm while releasing the fluoroscopy footswitch pedal. 2. Only perform these 2 actions sequentially. 3. If the system becomes locked up, perform a reset to recover functionalities. Customers were also informed that GE Healthcare will correct all affected products at n
Quantity49 devices
DistributionNationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122457GE INNOVA/INNOVA IGS/DISCOVERY IGS/OPTIMA ANGIOGRAPHIC, FLUOSCOPIC X-RAY SYSTEMS WITH CATHLAB FRONTIERS SOLUTIONS
GE Healthcare, Jan 2, 2013