Recall Z-1196-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Gmbh & Co. KG

Abbott Alinity i Processing Module LN 03R65-01 - Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

DateFeb 26, 2019
StatusTerminated
ReasonThe safety interlock covering the septum piercing probes within the bulk solution bottle holder may not deploy when a bulk solution bottle is removed from the Alinity ci series and Alinity s System.
Root causeUnder Investigation by firm
ActionOn February 26, 2019, Abbott sent a correction notice to there customers stating to please continue to follow the instructions provided in the Alinity ci series Operations Manual when replacing the bulk solutions. Use caution when accessing bulk solution bottle holders or replacing bulk solution bottles as you may be exposed to a sharp hazard. The Alinity ciseries Operations Manual will be updated in the future to add the appropriate hazard information to the procedure. If the safety interlock or bottle release does not function correctly, please contact your area customer support for assist
Quantity1097
DistributionWorldwide distribution - US Nationwide in the states of AR, CA, GA, ME, MN, MO, NY, OK, SC, TN, TX, UT, and VA, and countries of ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.