Recall Z-1137-2019 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

bioMerieux, Inc.

Biomerieux Vitek 2 Test kit VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

DateMar 15, 2019
StatusTerminated
ReasonUnexpected ESBL (Extended Spectrum Beta Lactamase) phenotype had been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N254, AST-N330, AST-N371 (ref.413723,418674, 422024) test kits.
Root causeSoftware change control
ActionOn 03/15/2019, the firm, BioMerieux sent notifications to all their consignees with the following instructions: "Please take the following actions at this time: Confirm this letter has been distributed to, and reviewed by, all appropriate personnel within your organization. Apply the VITEK 2 Systems bioART rule criteria as previously mentioned (i.e., using the criteria listed below for ESBL / E. coli isolates). Refer to Appendix A for detailed instructions. o If the card is AST-NXXX (where XXX indicates the card type designation, 254 / 330 / 371) o The organism is Escherichia coli o And the
Quantity10784 cartons
DistributionInternationally to: United Kingdom, Latvia, Poland, Ukraine, Germany, and Scotland.