Recall Z-1388-2019 ยท FSY Light, Surgical, Ceiling Mounted
Stryker Communications
EDS Light Suspension, Central Axis, a component of the Stryker Visum LED Surgical Lighting System, with the following catalog numbers: 0682001286, EDS Light Suspension, Central Axis, 5P/FP; 0682001298, EDS Light Suspension, Central Axis, 5P; 0682001299, EDS Light Suspension, Central Axis, 5P/5P; 0682001300, EDS Light Suspension, Central Axis, FP/5P/5P; 0682001432, EDS Light Suspension, Central Ax
| Date | Nov 13, 2018 |
| Status | Open, Classified |
| Reason | There is a potential the joint in the suspension of the device is insufficiently assembled. |
| Root cause | Employee error |
| Action | The recalling firm issued letters dated 11/13/2018 via FedEx informing to customers. The letter identified the affected product problem and actions to be taken. The customer is asked to disseminate the information within their organization. A response form was enclosed to indicate the customer has read and understands the instructions in the letter. The form was to be returned to the recalling firm via email. Customer was notified that a Stryker representative will begin visiting customer facilities in November to inspect their equipment. For questions contact " Stryker Technical Support |
| Quantity | 3,176 devices |
| Distribution | Distribution was nationwide, including Puerto Rico. There was also government/military distribution. Foreign distribution was made to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, France, India, Japan, Korea, Mexico, Netherlands, Spain and United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K060802 | VISUM LED SURGICAL LIGHTING SYSTEM Stryker Communications Corp., Apr 7, 2006 |