Recall Z-1130-2019 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

Philips North America, LLC

Azurion FlexArm, 722079

DateMar 12, 2019
StatusTerminated
ReasonDuring manufacturing of an Azurion R2.0 system, it was found out that the cable set was not according to specification. This resulted in damage of the 24V power cable of the detector.
Root causeComponent design/selection
ActionOn March 12, 2019, the firm distributed Urgent Medical Device Correction letters to affected consignees, alerting customers that the cable set in the C-arc of the Azurion R2.0 was not according to specifications, which resulted in damage of the 24V power cable of the detector. Customers were asked to place a copy of the field safety notice together with the documentation of the system until the system is inspected and corrected (if applicable) by Philips. The firm will inspect the affected systems free of charge and correct as necessary. If you need any further information or support concer
Quantity2
DistributionDistributed to one account in Florida, and one international account in Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181830Azurion R2.0
Philips Medical Systems Nederland B.V., Aug 2, 2018