Recall Z-1140-2019 ยท IZI System, X-Ray, Angiographic

Philips Medical Systems Nederlands

Foot Switches used with Philips MultiDiagnost Eleva with Flat Detector; Product Codes: 708037, 708038 Product The Allura Xper FD Series are intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures.

DateFeb 19, 2019
StatusTerminated
ReasonAdditional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
Root causeComponent design/selection
ActionThe firm previously notified customers on September 24, 2014 as part of Recall Z-1280-2016. On February 21, 2019, the firm sent an updated recall letter to newly identified affected consignees. Customers were alerted that bent pedal of footswitch might cause interruption of live Flouroscopy imaging or exposure. Action to be taken by Customer/User: 1. Before starting a procedure, the user should check the footswitch for possible bent pedals. If bent pedals are found the procedure should not continue and Local Service should be contacted. 2. Do not use the footswitch on an anti-fatigue mat or
Quantity7209 total
DistributionUS Nationwide Distribution in the states of CA, FL, MI, OR, PA, TX and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031333PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.2
Philips Medical Systems North America, Inc., May 13, 2003
K033737ALLURA XPER FD20
Philips Medical Systems North America, Inc., Dec 9, 2003
K041949ALLURA XPER FD10
Philips Medical Systems North America Co., Jul 30, 2004
K050151MULTIDIAGNOST ELEVA
Philips Medical Systems North America, Inc., Feb 9, 2005
K984545RELEASE 2 OF PHILIPS INTEGRIS SERIES (V5000, BV5000, H5000, BH5000, HM2000) SYSTEMS
Philips Medical Systems North America, Inc., Feb 25, 1999