Recall Z-1245-2019 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM Definition Edge, Model Number 10590000 Product Usage: Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

DateFeb 14, 2019
StatusTerminated
ReasonPotential risk of scans being aborted when using the optional Dual Spiral Dual Engery mode due to a software issue found in syngo CT VB10A for the Siemens Healthineers CT scanners with a single X-ray tube.
Root causeSoftware design
ActionSiemens issued letter dated 2/14/19 providing reason for recall and a user workaround in the Customer Safety Advisory Notice (CSAN) CT006/19/S. The recall issue will be permanently resolved by software update syngo CT VB10A_SP1. Contact the service organization 1-800-888-7436.
Quantity53 total units distributed in U.S.
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173630SOMATOM Force, SOMATOM Definition Flash, SOMATOM Drive, SOMATOM Definition Edge, SOMATOM Definition AS Open, SOMATOM Dei