Recall Z-1250-2019 ยท HGP Electrode, Circular (Spiral), Scalp And Applicator

Philips North America, LLC

Fetal Spiral Electrode, Model No. 9898 031 37631

DateJan 25, 2019
StatusTerminated
ReasonDuring use of the Philips FSE, it is possible for the metal electrode tip to break off and remain in the scalp of the newborn following labor and delivery.
Root causeUnder Investigation by firm
ActionOn January 25, 2019, the firm distributed Customer Information Medical Device Correction letters to affected customers. The letter informed customers that there was an increase in complaints regarding the Fetal Spiral Electrode tip breaking off, preventing easy removal from the scalp of the newborn infant. Customers were advised to do the following: * Visually examine the FSE upon removal from the newborn scalp to verify the tip is intact. If it is not intact, appropriate measures to locate and remove the missing part should be taken. * Upon receipt of this notice, clinicians should review t
Quantity1,971,925 (78,877 boxes)
DistributionDistributed nationwide within US. Worldwide international distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030691FETAL SPIRAL ELECTRODE, MODEL FSE07000
Clinical Innovations, Inc., Sep 17, 2003