Recall Z-1149-2019 ยท LFJ Catheter, Subclavian

Medical Components, Inc dba MedComp

14F SLX Double Lumen Full Tray Hemodialysis catheter insertion kits, Item No. MC061422

DateMar 13, 2019
StatusTerminated
ReasonTray label indicates the kits are packaged with a 15F "valved" peelable introducer, but the kits contain a 15F "non-valved" peelable introducer.
Root causeFinished device change control
ActionOn March 13, 2019, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed that the kits contain a 15F "non-valved" peelable introducer despite the label indicating that the kit is packaged with a "valved" introducer. Customers were asked to do the following: 1. Immediately examine your inventory and quarantine product subject to recall. 2. If you have further distributed this product, please identify your customers and immediately notify them of the recall. 3. Contact your Medical Components, Inc. customer service representative for a Returned Goods Auth
Quantity6830
DistributionDistributed to accounts in CA FL GA IN KY LA MA MI MO MS NJ NY OH OK PA SC TN TX VA WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K893439HEMO-CATH SILICONE DOUBLE LUMEN CATH SL28C & SL40C
Medical Components, Inc., Jun 13, 1989