Recall Z-1209-2019 ยท CAC Apparatus, Autotransfusion

Fenwal Inc

Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1. AT-1 sets and kits (9005104 AT-1 Autotransfusion set, 9108504 ATF 120 Fast Start Kit, and 9108494 ATF 40 Fast Start Kits); 2. AT-3 Autotransfusion sets and kits (Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits AT-3) Product Usage: The Fre

DateMar 5, 2019
StatusTerminated
ReasonBased on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the centrifuge tubing line making contact with the washing chamber. The contact may result in excessive wear of the centrifuge tubing or in extreme cases a puncture of the tubing.
Root causeDevice Design
ActionThe initial distributor was notified by Fresenius Kabi on 03/05/2019, who provided their distribution list. Fresenius Kabi notified those customers on about 03/11/2019 via "IMPORTANT - CATSmart Software Upgrade and AT-1 sets used on CATSmart" letter. Customers were instructed that Fresenius Kabi will be implementing a software changed to improve priming acceleration. As an interim solution, Fresenius Kabi instructed that CATSmart users should switch to using AT-3 Autotransfusion sets and kits: Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits (
Quantity342 devices
DistributionWorldwide distribution - US nationwide in the states of AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and PR, and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160735CATSmart
Fresenius Kabi AG, Sep 16, 2016