ClearanceBrief
Recall Z-1386-2019 ยท LGZ Warmer, Thermal, Infusion Fluid

Vyaire Medical

enFlow Disposable Cartridge with IV Extension Set, used for intravenous warming therapy with fluid and blood solutions.

DateJun 26, 2019 (initiated Mar 13, 2019)
StatusTerminated
ReasonTesting has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
Root causeDevice Design
ActionVyaire Medical notified consignees on about 03/13/2019 via URGENT NOTIFICATION OF RECALL letter. Consignees were instructed to conduct a 100% physical inventory to identify any affected product on hand, destroy all affected product in-stock in accordance with each facility's destruction protocol and if a facility is unable to destroy the product on site, or requires assistance, they were instructed to contact VyaireSupport@stericycle.com or call for assistance. Customers were also instructed to complete and return the Customer Response Form to VyaireSupport@stericycle.com, report any adverse
Quantity5,782,820 units total
DistributionWorldwide Distribution. US nationwide, Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Hungary, Israel, India, Jordan, Korea, Kuwait, Kazakhstan, Lebanon, Macedonia, Mexico, Malaysia , Myanmar, Netherlands, New Zealand, Oman, Peru, Qatar, Russian Federation, Saudi Arabia, Thailand, Turkey, Uni
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060537EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120
Enginivity, LLC, Jun 8, 2006
K130867ENFLOW IV FLUID WARMER INSULATED STRAP
Vital Signs, Inc., A GE Healthcare Company, Jun 14, 2013