Recall Z-1064-2019 ยท DZE Implant, Endosseous, Root-Form

Keystone Dental Inc

Paltop 1.25 Hex Drivers, Short Part Number: 60-70102

DateFeb 22, 2019
StatusTerminated
ReasonPaltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
Root causeProcess control
ActionKeystone Dental initiated recall by phone to each one of the impacted customers. The phone call was then followed by a Recall Notification Letter that was mailed via overnight FedEx mail on February 22, 2019. Letters states problem and action to take: Replacement 1.25mm Hex Driver(s) have been provided to you with this letter. The 1.25mm Hex Driver(s) currently in your kit should be discarded. Please refer to the images on page 2 to locate the driver within your kit. This issue is limited to this 1.25mm Hex Driver(s) only and no other product within the kit is affected by this issue. Complete
Quantity3 Hex Driver
DistributionFL, MA, SC, VA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K112795PALTOP DENTAL IMPLANT SYSTEM
Paltop Advanced Dental Solutions, Ltd., Apr 20, 2012