Recall Z-1133-2019 ยท PEN Gram-Negative Bacteria And Associated Resistance Markers
BioFire Diagnostics, LLC
FilmArray BCID Panel, IVD, Rx Only, BioFire Diagnostics, REF: RFIT-ASY-0126, and RFIT-ASY-0127, when used with the following blood culture bottles: Bottle Description/Part Number: BACT/ALERT FA Plus/410851; BACT/ALERT FN Plus/410852; BACT/ALERT PF Plus/410853
| Date | Mar 7, 2019 |
| Status | Terminated |
| Reason | Increased risk of false positive Enterobacteriaceae and Escherichia coli results on affected blood culture identification panels when using certain blood culture bottles. |
| Root cause | Material/Component Contamination |
| Action | On 03/07/19, Urgent: Medical Device Recall (Correction) notices were mailed to customers. Customers were informed that the false positive results are primarily Enterobacteriaceae results with a lower rate of concurrent E. coli results. If the affected BCID panel is used with BACT/ALERT blood culture bottles from the lots listed, then positive results for Escherichia coli and Enterobacteriaceae (if occurring without a specific genus/species reported other than E. coli; i.e. negative for all of the following: Enterobacter cloacae complex, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus, and |
| Quantity | 20350 |
| Distribution | MA, NH, VT, CT, NJ, NY, PA, DE, DC, VA, MD, WV, NC, SC, GA, FL, AL, TN, MS, KY, OH, IN, MI, IA, WI, MN, SD, ND, MT, IL, MO, KS, NE, LA, AR, OK, TX, CO, WY, ID, UT, AZ, NM, NV, CA, HI, OR, WA, AK, BM |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K130914 | FILMARRAY BLOOD CULTURE IDENTIFICATION (BCID) PANEL Biofire Diagnostics, Inc., Jun 21, 2013 |
| K143171 | FilmArray Blood Culture Identification (BCID) Panel for use with the FilmArray 2.0 Biofire Diagnostics, LLC, Jan 30, 2015 |
| K160457 | FilmArray Blood Culture Identification (BCID) panel for use with FilmArray Torch Biofire Diagnostics, LLC, Mar 15, 2016 |
| K181493 | FilmArray Blood Culture Identification (BCID) Panel Biofire Diagnostics, LLC, Jul 5, 2018 |