Recall Z-1128-2019 ยท LXH Orthopedic Manual Surgical Instrument

DePuy Orthopaedics, Inc.

CORAIL AMT COXA VARA NECK SEGMENT (KLA) Pro. Code L20432 Product Usage: The affected CORAIL Neck Trials are surgical instruments used in CORAIL total and partial hip arthroplasty.

DateSep 21, 2018
StatusTerminated
ReasonThere is the potential for debris/material to be found behind the O-rings in the neck trials.
Root causeNonconforming Material/Component
ActionOn September 21, 2018, the firm notified customers via email that rework was required for the affected product. Customers were advised that the three affected product codes were produced with an O-Ring until a 2010 design change removed the O-Ring as a rolling change. Customers were provided with work instructions for how to rework the product by removing the O-Ring. Please Take the Following Urgent Actions: 1. Please continue to follow the instructions for use in IFU-W90946 Rev B regarding cleaning of these devices. 2. Immediately review your inventory and rework the affected units. Your De
Quantity12639 total
DistributionWorldwide Distribution - US Nationwide Distributed to accounts in AL AZ CA CO FL GA HI IA IL IN KY LA MA MD ME MI MN MS NC ND NH NJ NV NY OH OR PA SC TN TX UT VA WA, and WI. International distribution worldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.