Recall Z-0252-2020 ยท FMI Needle, Hypodermic, Single Lumen

Becton Dickinson & Company

BD Vacutainer Eclipse Blood Collection Needles 22Gx1.25, cat. no. 368608

DateOct 31, 2019 (initiated Mar 7, 2019)
StatusTerminated
ReasonBD PAS received customer complaints indicating the absence of the bevel on the non-patient needle end of the product, causing blood leakage.
Root causeNonconforming Material/Component
ActionOn March 7, 2019, the firm distributed Urgent Medical Device Recall letters to customers. Customers were informed of the missing bevel, which may ultimately cause leakage if the needle damages the sleeve. ***UPDATED 7/11/19*** In July 2019, the firm sent out an additional notice, informing customers that two additional lots of the product were found to also be affected by the recall. ***UPDATED 12/10/19*** ON 12/4/19, the firm sent out a second expansion letter, informing that customers that an additional lot of product (lot 9010765) was also found to be missing the bevel from the non-pati
Quantity2,624,160 (2,169,703 US; 454,457 OUS)
DistributionDistributed to accounts in AR, CT, FL, IL, IN, MI, MO, MS, NY, OH, TN, TX, and VA. Foreign distribution to Canada, Brazil, India, Mexico, Singapore, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982541VACUTAINER BRAND ECLIPSE BLOOD COLLECTION NEEDLE, MODELS# 368607 & 368608
Becton Dickinson Vacutainer Systems, Oct 28, 1998