Recall Z-1373-2019 ยท CGA Glucose Oxidase, Glucose

Polymer Technology Systems, Inc.

CHOL+HDL+GLU test strips, REF 2412

DateJan 16, 2019
StatusTerminated
ReasonSome lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed
Root causeDevice Design
ActionThe firm initially notified some of their customers from January through March 2019 with a Customer Bulletin. On March 28, 2019, the firm issued an updated Urgent Medical Device Recall letter to all affected customers. The letter alerted customers to the potential for test strips to slip out of the analyzer. Actions to be taken by the Customer/User: 1. Closely follow the instructions for proper testing. " Make sure the analyzer is sitting on a flat, stable surface. Do not perform testing with the analyzer held in the hand. " After inserting the test strip(s), do not move the analyzer unt
Quantity67643 total
DistributionDomestic distribution nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071507PTS PANELS CHOL+HDL+GLU PANEL TEST STRIPS
Polymer Technology Systems, Inc., Sep 10, 2007
K182781PTS Professional Chemistry Kit
Polymer Technology Systems, Inc. D/B/A Pts Diagnostics, Dec 21, 2018