Recall Z-1138-2019 ยท NFJ System, Image Management, Ophthalmic

Topcon Medical Systems, Inc.

IMAGEnet 6 v1.53- IMAGEnet 6 Ophthalmic Data System is a software program that is intended for use in the collection, storage and management of digital images, patient data, diagnostic data and clinical information from Topcon devices without controlling or altering the functions and parameters of any medical devices or through computerized networks.

DateApr 12, 2019 (initiated Mar 6, 2019)
StatusTerminated
ReasonWhen a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.
Root causeLabeling design
ActionTopcon issued Device Correction Notice on 3/22/19 providing problem- there exists a situation in which the displayed patient name and the thumbnail image do not match and action to take.Topcon has corrected this issue in IMAGEnet 6 version 2.0.3 and will be providing a remote upgrade free of charge. Contact customer support team to schedule the upgrade at 866.922.6278 option 1 or by email at TMSservice@topcon.com
Quantity450 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171370IMAGEnet 6 Ophthalmic Data System
Topcon Corporation, Nov 1, 2017