Recall Z-1122-2019 ยท JKA Tubes, Vials, Systems, Serum Separators, Blood Collection

Becton Dickinson & Company

BD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010

DateMar 7, 2019
StatusTerminated
ReasonThe luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.
Root causeProcess change control
ActionUrgent Medical Device Recall notification letters dated 3/7/19 were distributed to customers.
Quantity6,509,600
DistributionUS distribution to the states of FL and NJ.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972404VACUTAINER BRAND SAFETY BLOOD COLLECTION ASSEMBLY(MULTIPLE)
Bd Becton Dickinson Vacutainer Systems Preanalytic, Jul 22, 1997