Recall Z-1053-2019 ยท IXR Table, Radiographic, Tilting

Philips North America, LLC

Urodiagnost, Product No. 708033 Product Usage: The Philips UroDiagnost MRF can be used to perform all urological and gynecological routine examinations including: Urograms with injection or infusion Retrograde pyelography Cystography and urethrography Micturition cystography Cystoscopy, endopscopy Transurethral resection Percutaneous nephroscopy Urethrorenoscopy Percutaneous nep

DateFeb 28, 2019
StatusTerminated
ReasonThe locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
Root causeComponent design/selection
ActionOn February 28, 2019, the firm notified customers via an Urgent Field Safety Notice letter. The letter informed customers that the affected systems would be inspected for missing locking plates. Customers were asked to place a copy of the Field Safety Notice with the documentation of the system until the system has been inspected by Philips. Philips will inspect all potentially affected systems free of charge. If a locking plate is found missing, Philips will secure the system until a locking plate is installed so that use of the system can continue. You will be contacted by your local Phili
Quantity1969
DistributionUS Nationwide domestic distribution, and worldwide international distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911098PHILIPS URO DIAGNOST MRF
Philips Medical Systems, Inc., Jul 3, 1991