Recall Z-1292-2019 ยท KHO Fluorometer, For Clinical Use

Ortho-Clinical Diagnostics

VITROS ECi/ECiQ Immunodiagnostic System Product Code: 1922814 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents

DateFeb 20, 2019
StatusTerminated
ReasonLuminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS Systems
Root causeSoftware design
ActionOrtho Clinical Diagnostics (Ortho) issued an "URGENT PRODUCT CORRECTION NOTIFICATION" , a customer letter (Ref. CL2019-052) dated 21 February 2019 via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to customers who were shipped affected VITROS Systems/or whose VITROS System has a luminometer parts replacement. The customers were notified of the issue and instructed to notify Ortho upon experiencing the issue described: -Avoid performing a shutdown/restart on your systems unless directed by an Ortho representative or prompted by condition code help text. -I
Quantity45 units
DistributionWorldwide Distribution: US (Nationwide) and countries of: Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden,The Netherlands, and United Kingdom.