ClearanceBrief
Recall Z-1287-2019 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Ortho-Clinical Diagnostics

VITROS 5600 Integrated System -Software V3.3.2 & below Product Code: 6802413 For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Immunodiagnostic Products Reagents

DateFeb 20, 2019
StatusTerminated
ReasonLuminometer Malfunction May Cause Inability to Process MicroWell Assays on VITROS Systems
Root causeSoftware design
ActionOrtho Clinical Diagnostics (Ortho) issued an "URGENT PRODUCT CORRECTION NOTIFICATION" , a customer letter (Ref. CL2019-052) dated 21 February 2019 via FedEx overnight courier and/or US Postal Service Priority Mail (for PO Boxes only) to customers who were shipped affected VITROS Systems/or whose VITROS System has a luminometer parts replacement. The customers were notified of the issue and instructed to notify Ortho upon experiencing the issue described: -Avoid performing a shutdown/restart on your systems unless directed by an Ortho representative or prompted by condition code help text. -I
Quantity55 units
DistributionWorldwide Distribution: US (Nationwide) and countries of: Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden,The Netherlands, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081543VITROS IMMUNODIAGNOSTIC PRODUCTS TSH REAGENT PACK AND CALIBRATORS, VITROS 5600 INTEGRATED SYSTEM
Ortho-Clinical Diagnostics, Inc., Oct 23, 2008