Recall Z-1210-2019 ยท FPA Set, Administration, Intravascular

Becton Dickinson & Company

SmartSite Syringe Administration Set-Product Package Size 50, Catalog Number: 10798696

DateJun 26, 2019 (initiated Feb 28, 2019)
StatusTerminated
ReasonLeaking of the Smartsite Syringe Administration Set. May result in delay or interruption of infusion, under-infusion, contamination of the fluid path and HCP exposure to infusates.
Root causeProcess control
ActionBD distributors provided a hard copy of the recall letter on February 28, 2019 via FedEx and may also be contacted via e-mail (when available). Distributors are requested to identify their customers and provide BD a customer list so that BD can undertake further communication to the users. BD will distribute the recall communication to the distributor customers via FedEx upon receiving the customer information from the distributors. A copy of the BD product recall customer letter was provided to distributors, in the case they opt to contact their customers directly. BD is requesting that cust
Quantity2900 Sets
DistributionUS Nationwide Distribution in the states of AR, NH, NC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K81188530-30-2 700 SYRINGE EXTENSION SET
Y, Aug 25, 1981