Recall Z-1032-2019 ยท FTM Mesh, Surgical

ACell, Inc

Gentrix Surgical Matrix Thick, 30 x 40cm, Catalog Number PSMT3040, Lot 013948 - Product Usage: Gentrix Surgical Matrix Thick is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic and reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limi

DateFeb 14, 2019
StatusTerminated
ReasonThe devices were released from a lot for which one sample failed to meet the company' s tensile strength specification for lot release.
Root causeNonconforming Material/Component
ActionUrgent Medical Device Recall notification letters dated 2/14/19 were sent to customers.
Quantity1
DistributionUS nationwide distribution in the state of CA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170763Gentrix Surgical Matrix Thick; Gentrix Surgical Matrix Extend
Acell, Inc., Jun 8, 2017