Recall Z-0985-2019 ยท OWB Interventional Fluoroscopic X-Ray System
Siemens Medical Solutions USA, Inc
Uroskop Omnia Max -fluoroscopic X-ray system Model # 10762473 Uroskop Omnia Max is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
| Date | Feb 8, 2019 |
| Status | Terminated |
| Reason | The swivel arm holding the monitors may become dislodged from the carrying arm. |
| Root cause | Device Design |
| Action | The firm, Siemens, issued an "URGENT: MEDICAL DEVICE CORRECTION "on 2/8/19 to its customers. The letter is identifying the problem, risk, actions to take and corrective action. The customers were instructed to do the following: Immediately examine your inventory and quarantine product subject to recall (if applicable). or To avoid the risk until implementation of the corrective action, it is strongly recommended not to move and position monitors above the system table. As an alternative and if clinically applicable, monitors can also be used when they are left in the parking position. In addit |
| Quantity | 191 |
| Distribution | US Nationwide Distribution |