ClearanceBrief
Recall Z-1076-2019 ยท CGS Nad Reduction/Nadh Oxidation, Cpk Or Isoenzymes

Abbott Laboratories, Inc

Abbott ARCHITECT Creatine Kinase, IVD, REF 7D6341.

DateJul 18, 2018
StatusTerminated
ReasonThe product has a stability issue which may lead to an error code indicating "Unable to calculate result, reaction check failure" and/or Quality Control results out of range low.
Root causeMaterial/Component Contamination
ActionRecall letters were issued via FedEx overnight priority mail on 7/18/2018.
Quantity30 kits
DistributionDistribution was made nationwide, including Puerto Rico. There was also government and military distribution. Foreign distribution was made to Algeria, Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983070CK
Abbott Laboratories, Feb 16, 1999