Recall Z-1040-2019 ยท IZL System, X-Ray, Mobile

Fujifilm Medical Systems U.S.A., Inc.

Fujifilm FDR Go PLUS, Mobile X-Ray System Product Usage: This device is intended to be used to take a film, CR or general radiography of patients in the hospital who cannot move and/or of outpatients in emergency. Exposure can be captured on film, CR or Digital Radiography (DR)>

DateJan 15, 2019
StatusTerminated
ReasonFUJIFILM Medical Systems U.S.A., Inc., Modality Systems became aware of the possibility that the braking mechanism might not be engaged in all cases when the driving handle has been released.
Root causeDevice Design
ActionFujifilm sent an Urgent Medical Device Correction and Removal letter dated January 16, 2019 to customers. The letter identified the affected product, problem and actions to be taken. Customers were advised to schedule a visit by our Service department to replace the handle assembly. Customers were provided with additional instructions to continue using the affected product safely until the correction was made. For questions call 201-675-8947.
Quantity56 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080701SHIMADZU MOBILE X-RAY SYSTEM MOBILEDART EVOLUTION
Shimadzu Corp., May 14, 2008